Skip to main content
All CollectionsRPM
Troubleshooting an RPM Blood Pressure Monitor
Troubleshooting an RPM Blood Pressure Monitor

How to Resolve Common Problems and Error Codes

Bill Ruby avatar
Written by Bill Ruby
Updated over 4 years ago

Below are some common troubleshooting issues and how to resolve them:

Problem: Device doesn’t power on when START/STOP button is pressed. or Cuff does not inflate. (display remains off).

  • Possible Reason: Batteries completely drained.
    Resolution:
    Replace batteries with new Duracell AA.

  • Possible Reason: Batteries inserted with incorrect polarity.
    Resolution:
    Remove batteries and reinstall them making sure the + and - polarity symbols on the batteries match those inside the battery compartment.

Problem: Cuff does not inflate. (display turns on).

  • Possible Reason: Cuff not plugged in correct.
    Resolution:
    Make sure the connector on the cuff tube is plugged all the way into the device.

Problem: Cuff doesn’t form a loop around arm.

  • Possible Reason: Cuff isn’t assembled properly.
    Resolution:
    To form a loop: pass the end of the cuff furthest away from the tubing through the long metal D-ring. The smooth side (without the hook and loop fasteners) should be on the inside of the cuff loop.

Problem: Readings too high or low OR readings vary significantly between measurements.

  • Possible Reason: Not using proper cuff size
    Resolution:
    Your BP Monitor comes with a standard, one-size-fits-all cuff that is 22 - 42 cm. If you are finding that the one-size-fits-all cuff is too loose, avoid wrapping the cuff around your arm multiple times as this can give inaccurate readings. Instead, you can order a smaller replacement cuff. The smaller cuff is 22 - 32 cm, and works better for people with smaller arms. Contact ThoroughCare about ordering your smaller replacement cuff.

  • Possible Reason: Cuff not positioned correct.
    Resolution:
    Review the instructions on Securing the Cuff, and retry measurement. Specifically make sure the cuff is not to loose or too tight.

  • Possible Reason: Movement during measurement.
    Resolution:
    Place your arm on a table with your palm facing up. The cuff should be at the same level as your heart. Make sure you stay still and don’t talk during measurement.

Problem: Measurement results differ from those measured at doctor’s office.

  • Possible Reason: Measurements taken in a medical/professional environment (white coat effect) may affect your blood pressure.
    Resolution:
    Measurements taken in a more relaxed, home environment may avoid effects experienced in clinical settings. Continue taking readings as usual. If in doubt, contact a medical professional.

Problem: Error code shown after measurement.

  • Possible Reason: Measurement accuracy may be affected or transmission may have been unsuccessful.
    Resolution:
    Consult the error code table below for more information and further instructions.

Error codes

Code: Er1
Description:
Low battery error.
Resolution:
Confirm that the replacement batteries are all brand new, unexpired, Duracell branded. Confirm that the battery compartment is clean and the contact terminals have no foreign material blocking them.

Code: Er2
Description:
Movement detected.
Resolution:
Avoid moving or talking while taking a measurement. After securing the cuff assuming the correct posture, consider waiting for five minutes to settle down before starting a measurement.

Code: Er3
Description:
Measurement out of range.
Resolution:
Confirm that the cuff is secured correctly on your exposed arm. Remove all clothing and jewelry from your arm. If you feel unwell, contact a medical professional.

Code: Er4 (batt)
Description:
Low battery error.
Resolution:
Confirm that the replacement batteries are all brand new, unexpired, Duracell branded. Confirm that the battery compartment is clean and the contact terminals have no foreign material blocking them.

Code: Er4 (cell)
Description:
cell/network error (displayed after lines going around).
Resolution:
Confirm test at EXACT SAME location. Confirm text at NEW location, preferably closer to a window. Take three successive measurements.

Code: Er5
Description:
cellular connection error (displayed after lines going around); network found, unable to set-up data connection.
Resolution:
Confirm test at EXACT SAME location. Confirm text at NEW location, preferably closer to a window. Take three successive measurements.

Code: Er6
Description:
Data transmission error (displayed after lines going around); signal not strong enough.
Resolution:
Confirm test at EXACT SAME location. Confirm text at NEW location, preferably closer to a window. Take three successive measurements.

Code: Er7
Description:
Internal error (displayed after lines going around).
Resolution:
Check that replacement batteries are all brand new, unexpired, Duracell branded.

Code: Er9
Description:
Provisioning error (displayed after lines going around).
Resolution:
Request assistance from ThoroughCare.

Code: DISPLAY
Description:
Display issues: completely blank, segment missing, incorrect characters.
Resolution:
Confirm the participant understands power-on procedure. Check that replacement batteries are all brand new, unexpired, Duracell branded.

Code: PHYS
Description:
Physical defects: cracked glass or plastic, large scratches, missing feet, etc. impacting performance.
Resolution:
Confirm that the defect impedes the function of the device.

Code: WAKE
Description:
Unit won’t power on.
Resolution:
Review and confirm that you understand the power-on procedure. Check that replacement batteries are all brand new, unexpired, Duracell branded.

Code: LOST
Description:
Product lost in shipping.
Resolution:
Contact ThoroughCare.

Code: Er10
Description:
Subscription suspended (displayed after lines going around).
Resolution:
Contact ThoroughCare.

Code: Er11
Description:
Terminated (displayed after lines going around).
Resolution:
Permanently deactivated (requested by customer).

Code: Er15
Description:
Calibration error.
Resolution:
Contact ThoroughCare.

Code: Er16
Description:
Internal memory error.
Resolution:
Contact ThoroughCare.

Code: Er17
Description:
Internal memory corruption.
Resolution:
Contact ThoroughCare.

Code: Er18
Description:
Sensor error.
Resolution:
Contact ThoroughCare.

Code: Er19
Description:
Failed to zero.
Resolution:
Contact ThoroughCare.

Code: Er20
Description:
Cuff didn’t deflate.
Resolution:
Check cuff placement. Make sure the connector is plugged in and the tube is straight.

Code: Er21
Description:
Pulse not detected.
Resolution:
Check cuff placement. Make sure the connector is plugged in and the tube is straight.

Code: Er22
Description:
Measurement error.
Resolution:
Check cuff placement. Make sure the connector is plugged in and the tube is straight.

Code: Er23
Description:
Hardware failure.
Resolution:
Contact ThoroughCare.

Code: Er24
Description:
Software failure.
Resolution:
Contact ThoroughCare.

Did this answer your question?